Oval gear flowmeter displacement meter, also known as fixed displacement meter, elliptical gear at the beginning of each rotation 4 times a week can be output volume of the cavity, therefore, the elliptical gear revolution is proportional to the fluid flow.
Fault | Reason | Treatment methods |
elliptic gear does not rotate | 1.debris in the pipe | 1.Clean the flowmeter and pipe. |
2.filter is damaged, the sundries are in the table, and the gear is stuck.; | 2.Repair filters; | |
3.Because of the excessive flow, bearing wear, the gear collides with the bore, causing the gear to jam. | 3.Replace the bearing to ensure the clearance is flexible. | |
Elliptical gear rotation Pointer does not move | 4.Rotating fasteners sliding, damaged or not flexible; | 4.Tighten the fastener to remove the inflexible; |
5.Magnetic coupling sliding; | 5.Check whether the magnetic coupling is flexible and strong; | |
6.The word wheel part is stuck. | 6.Troubleshoot the word wheel parts. | |
Flow meter has a large positive error | 7.The flow has a large pulsation or contains gas; | 7.Reduce flow pulsation or increase gas separator; |
8.The actual liquid viscosity is greater than the test liquid viscosity; | 8.Select the test liquid to check the flow meter or viscosity correction with the actual liquid viscosity; | |
9.Small inlet valve, medium not full of measurement room measurement. | 9.Use flow meter outlet valve to control medium flow. |
The flowmeter are normally possible fault can be found in the above account causes and corresponding processing methods, if a fault does not find a reason, you can contact us, providing professional solutions.